Purpose. At low q, ln I vs q² is approximately linear for a compact particle. The slope gives the radius of gyration Rg; the intercept gives I(0). The chosen interval also defines where later tools (GNOM, BODIES) should start reading the curve.
What autosaxs does (custom) — adaptive Guinier. Rather than trusting a single fixed window or ATSAS AUTORG alone, autosaxs:
In liveview, open the Guinier Adjust wizard to override first / last (1-based point indices) for a fixed-interval fit; confirming suspends auto-processing until you resume. The main Guinier pane shows a Passport table (not inline spin boxes).
How to read the UI. The passport summarizes fit quality, low-q classification, Rg, q-range, and related diagnostics for the selected interval. Fit quality is one of: good, acceptable, weak in Guinier region, weak, degenerate. Low-q classification is one of: linear, upturn, downturn, chaotic. A “chaotic” or poor fit-quality means the low-q region is unreliable for Rg even if a number is shown — treat GNOM and shape steps with caution. Passport rows may show a FAILED / warning status when quality checks do not pass.
Engines. Custom adaptive search (+ optional ATSAS AUTORG for comparison).