Masking
In order to mask the non-test ear (NTE) during air-conduction audiometry set the Chan. 2 level
to the desired value and make sure the Chan. 2 ON
checkbox is checked.
For masking during bone conduction audiometry, additionnally, the Non-test ear status
has to be set to Earphone on
(otherwise the software considers that the noise is being played through the earphone, but the earphone is not on the listener’s ear, so no noise actually arrives at the listener’s ear). audiometry_trainer
automatically sets the Non-test ear status
to Earphone on
when you check the Chan. 2 ON
checkbox. This small complication for masking in bone conduction audiometry is due to the fact that audiometry_trainer
lets you test bone conduction thresholds either with the NTE uncovered, or with the NTE covered by the earphone while the earphone is not playing any noise. This is useful to measure the size of the occlusion effect (see Measuring the occlusion effect).
The Lock channels
checkbox can be used to “lock” the test ear (TE) and NTE channels so that changing the signal level sent to the TE changes the noise level sent to the NTE by the same amount.
Interaural attenuation
Interaural attenuation (IA) values for supra-aural and insert earphones are drawn randomly for each case from a uniform distribution. The upper and lower limits of the distribution are frequency specific and are based on the max and min IA values reported by [MunroAndAgnew1999]. For frquencies that were not tested in [MunroAndAgnew1999] the values of an adjacent frequency are used.
IA values for bone conduction (BC) are drawn for each case from a uniform distribution. The lower and upper values for the uniform distribution as a function of frequency is shown in Table table-IA_bone. These values were chosen to reflect the observation that IA for bone conducted stimuli using a mastoid placement is essentially zero at 250 Hz and increases up to ~15 dB at 4000 Hz ([Studebaker1967]; [Gelfand2016]). More nuanced data on IA for BC are available [Stenfelt2012] and may be integrated to future versions of audiometry_trainer
.
Freq. |
Low |
High |
---|---|---|
125 Hz |
0 |
0 |
250 Hz |
0 |
0 |
500 Hz |
0 |
2 |
750 Hz |
0 |
3 |
1000 Hz |
0 |
4 |
1500 Hz |
0 |
5 |
2000 Hz |
0 |
7 |
3000 Hz |
0 |
10 |
4000 Hz |
0 |
15 |
6000 Hz |
0 |
15 |
8000 Hz |
0 |
15 |
Occlusion effect
For cases without a conductive hearing loss the size of the occlusion effect (OE) is drawn randomly for each case and frequency from a uniform distribution. The lower and upper values for the uniform distribution as a function of frequency is shown in Table table-OE_supra for supra-aural headphones and Table table-OE_insert for insert earphones. These values were chosen to reflect approximately the range of values found in various studies and summarized in [Gelfand2016].
Freq. |
Low |
High |
---|---|---|
125 Hz |
15 |
30 |
125 Hz |
15 |
30 |
500 Hz |
8 |
26 |
750 Hz |
8 |
26 |
1000 Hz |
4 |
12 |
Freq. |
Low |
High |
---|---|---|
125 Hz |
2 |
10 |
125 Hz |
2 |
10 |
500 Hz |
2 |
10 |
750 Hz |
2 |
10 |
1000 Hz |
2 |
10 |
The OE is absent when a conductive/mixed hearing with an air-bone gap (ABG) >= 20 dB is present [MartinEtAl1974]. For this reason, if there is an ABG >= 20 dB the OE is set to zero. For ABGs between 0 and 20 dB the size of the OE is scaled by the ABG using the following equation:
where \(OE_{inp}\) is the OE before the scaling and \(OE_{out}\) is the OE after the scaling. This means that the size of the OE is progressively reduced as the ABG increases, reaching zero for an ABG of 20 dB.
Measuring the occlusion effect
Some authors (e.g. [Gelfand2016]) recommend measuring the OE in individual listeners to determine the value to be used in masking formulas. This can be done by covering the NTE with the earphone and measuring BC thresholds, without presenting any masking noise to the NTE. The thresholds obtained in this “earphones ON” condition can then be compared to those obtained without covering the NTE to estimate the magnitude of the occlusion effect [MartinEtAl1974]. To obtain BC thresholds with the NTE covered by an earphone, but without delivery of masking noise, in audiometry_trainer
you need to set the Non-test ear status
to Earphones ON
and make sure that the Chan. 2 ON
checkbox is uncheked. The size of the OE measured in this way will depend on the earphones used for Chan. 2
, so make sure that you select the same earphone type that will be later used to deliver the masking noise.
A limitation of this technique is that in some cases it can underestimate the size of the OE ([FagelsonAndMartin1994]; [Yacullo1997]). This may happen in the case of a listener with NTE BC thresholds close to the lower output limits of the BC vibrator. For example, if the listener has a BC threshold of 0 dB HL and the BC vibrator has a lower output level limit of -20 dB HL, then OEs larger than 20 dB would be underestimated. This may also happen if the TE BC threshold is lower than the NTE BC threshold. Suppose for example that the TE BC threshold is 20 dB HL, the NTE BC threshold is 30 dB HL, and the NTE OE is 30 dB. The threshold recorded in the unoccluded condition will be 20 dB HL and when the NTE is occluded the threshold recorded will be 0 dB HL; the difference between the two, the estimated OE, will be 20 dB rather than the actual NTE OE of 30 dB. If masking is used only to confirm that the unmasked TE BC threshold is genuine, underestimating the OE will be of no consequence; the TE had the lowest threshold to start with, so undermasking will simply confirm, “in the wrong way”, a correct decision that that threshold was coming from the TE. If the estimated OE is used for a full-blown masked threshold search, the masked TE BC threshold may be lower than the unmasked one because the NTE will be undermasked.