Argus II: images, video, and encoders
Argus II converts each camera frame into biphasic pulses on 60 epiretinal electrodes. This example follows an image and a video through that chain:
and changes one link at a time: the model, the encoder, and the source.
import matplotlib.pyplot as plt
import numpy as np
from pulse2percept.implants.retina import ArgusII
from pulse2percept.models.retina import AxonMapModel, BiphasicAxonMapModel
from pulse2percept.models.retina import ScoreboardModel
from pulse2percept.stimuli import (AmplitudeEncoder, FrequencyEncoder,
psychophysics, samples)
from pulse2percept.units import Hz, dva, s, uA
From image to delivered stimulation
ArgusII() carries its device defaults: an
AmplitudeEncoder that maps gray levels
0-1 to 0-50 uA at 6 Hz, and a
SequentialRaster that pulses one row at
a time. The image is stretched across the array, so each electrode receives
the gray level at its own position:
implant = ArgusII()
image = samples.logo_ucsb()
fig, (ax_src, ax_dev) = plt.subplots(ncols=2, figsize=(11, 3.5))
image.plot(ax=ax_src)
ax_src.set_title('Source')
implant.plot(stim=image, stim_cmap=True, ax=ax_dev)
ax_dev.set_title('Delivered amplitude per electrode')
fig.tight_layout()

Same stimulation, two models
The model sets how the delivered pulses appear.
ScoreboardModel draws one round
phosphene per electrode;
AxonMapModel spreads each one along
the nerve fiber bundles passing under it [Beyeler2019]. rho (um) is the
spread across bundles and lam (um) the spread along them:
models = {
'ScoreboardModel': ScoreboardModel(implant=implant, rho=100),
'AxonMapModel': AxonMapModel(implant=implant, rho=100, lam=300),
}
fig, axes = plt.subplots(ncols=2, figsize=(10, 4))
for ax, (name, model) in zip(axes, models.items()):
model.predict_percept(image).plot(ax=ax)
ax.set_title(name)
fig.tight_layout()

Choosing an encoder
Spatial-only models see one gray level per electrode, so the encoder’s pulse
parameters do not change their output. A spatiotemporal model reads the
delivered pulse train.
BiphasicAxonMapModel [Granley2021]
lets amplitude scale phosphene size and brightness, while frequency scales
brightness only.
The model takes amplitude as a multiple of perceptual threshold, so every electrode is given an assumed 50 uA threshold. Both encoders below drive a white pixel at the same charge per second (150 uA x 20 Hz = 50 uA x 60 Hz):
implant.thresholds = 50 * uA
model = BiphasicAxonMapModel(implant=implant, rho=100, lam=300)
encoders = {
'Amplitude: 0-150 uA at 20 Hz':
AmplitudeEncoder(amp_range=(0, 150) * uA, freq=20 * Hz),
'Frequency: 0-60 Hz at 50 uA':
FrequencyEncoder(amp=50 * uA, freq_range=(0, 60) * Hz),
}
percepts = {}
for label, encoder in encoders.items():
implant.encoder = encoder
percepts[label] = model.predict_percept(image)
vmax = max(p.data.max() for p in percepts.values())
fig, axes = plt.subplots(ncols=2, figsize=(10, 4))
for ax, (label, percept) in zip(axes, percepts.items()):
percept.plot(ax=ax, vmin=0, vmax=vmax)
ax.set_title(label)
fig.tight_layout()

Both panels share one brightness scale. Amplitude encoding recruits wider phosphenes; frequency encoding keeps them narrow.
Video
A moving bar is a common Argus II motion task. The bar is defined in the
visual field: a Scene 30 dva wide, sampled
every 50 ms. Each electrode receives the part of the scene at its own
visual-field location (see Visual Input and Simulated Vision).
The default 6 Hz pulse rate is slower than the 20 fps video, so most frames would deliver no pulse. A 20 Hz encoder delivers one pulse per frame:
implant = ArgusII()
implant.encoder = AmplitudeEncoder(freq=20 * Hz)
model = AxonMapModel(implant=implant, rho=100, lam=300)
bar = psychophysics.bar(width=2 * dva, speed=20 * dva / s, offset=-12 * dva,
fov=30 * dva, time=np.arange(0, 1200, 50))
percept = model.predict_percept(bar, gaze=(0, 0) * dva)
percept.play()
What this does not establish
rho,lamand the 50 uA threshold are illustrative values. Measured values vary across subjects and electrodes.Model brightness is in arbitrary units.
Argus II’s camera-to-electrode mapping is configured per patient and is not modeled; each electrode samples its own visual-field location.