20240830
satrec  T1
optimized echo
calibraitng pulse length - 20220406
17O 1174 Hz (spinning frequency)
rotor LAMOX 17O with a little bit KBr and BN, 1 mm sample layer, plus BN packed layer by layer up to the brim of the can
p1=5 us at 500 W (from d1=60s popt 998 expno 14)

Laser Current : 25 A
cnst31=20kHz
de=45us

PLW1 500W

Amplifier control still unbalanced at 300W. And peak power of the probe is around 650W. So I will revert to 500W as the probe cant be damaged at this PLW1. However, the question remains if peak power of a probe
changes with usage.

ns =

d1=0.2 s

shortening the vdlist and trying way less time

VD_LIST:

0.000005
0.00001
0.00003
0.00005
0.0001
0.0003
0.0005
0.001
0.003
0.005
0.01
0.03
0.05
0.1
0.3