Metadata-Version: 2.4
Name: eyelink1000plus-pupil-size-to-mm
Version: 1.1.0
Summary: EyeLink 1000 Plus pupil-size calibration (arbitrary units → mm) via an artificial-eye recording.
Project-URL: Homepage, https://github.com/mh-salari/eyelink1000plus-pupil-size-to-mm
Project-URL: Repository, https://github.com/mh-salari/eyelink1000plus-pupil-size-to-mm
Project-URL: Issues, https://github.com/mh-salari/eyelink1000plus-pupil-size-to-mm/issues
Author-email: Mohammadhossein Salari <mohammadhossein.salari@gmail.com>
License: MIT
License-File: LICENSE
Keywords: calibration,eyelink,eyelink-1000-plus,eyetracking,neuroscience,pupil-size,pupillometry
Classifier: Development Status :: 3 - Alpha
Classifier: Intended Audience :: Science/Research
Classifier: License :: OSI Approved :: MIT License
Classifier: Programming Language :: Python :: 3
Classifier: Programming Language :: Python :: 3.12
Classifier: Topic :: Scientific/Engineering
Requires-Python: <3.13,>=3.12
Requires-Dist: pyelink[pyglet]>=1.1.5
Requires-Dist: sr-research-pylink>=2.1.1145.post1
Provides-Extra: dev
Requires-Dist: pytest; extra == 'dev'
Requires-Dist: ruff>=0.15.13; extra == 'dev'
Description-Content-Type: text/markdown

# eyelink1000plus-pupil-size-to-mm

[![PyPI version](https://img.shields.io/pypi/v/eyelink1000plus-pupil-size-to-mm)](https://pypi.org/project/eyelink1000plus-pupil-size-to-mm/)
[![Downloads](https://static.pepy.tech/badge/eyelink1000plus-pupil-size-to-mm)](https://pepy.tech/project/eyelink1000plus-pupil-size-to-mm)
[![License](https://img.shields.io/pypi/l/eyelink1000plus-pupil-size-to-mm)](https://github.com/mh-salari/eyelink1000plus-pupil-size-to-mm/blob/main/LICENSE)
[![DOI](https://zenodo.org/badge/1242921854.svg)](https://doi.org/10.5281/zenodo.20277434)

Convert EyeLink 1000 Plus pupil-size readings (arbitrary units) into millimetres.

The EyeLink 1000 Plus records pupil size in arbitrary units whose scale depends on the camera distance and the tracking threshold the Host PC chose for that recording. To report pupil size in physical units, you record a printed circle of known diameter (an *artificial eye*) at the camera-to-eye distance your participants will sit at, derive a per-eye unit-to-mm constant from that recording, and apply the constant to every participant recording made with the same physical setup. The procedure follows SR Research's [FAQ How can I convert pupil size to mm?](docs/FAQ%20How%20can%20I%20convert%20pupil%20size%20to%20mm%3F.md) (a copy is shipped under `docs/`).

Two conversion formulas are supported, matching the Host PC's `pupil_size_diameter` setting at recording time:

| Host PC mode                | Formula                            |
|-----------------------------|------------------------------------|
| `pupil_size_diameter = NO`  | `mm = sqrt(units) / constant`      |
| `pupil_size_diameter = YES` | `mm = units / constant`            |

The Host PC mode used for the artificial-eye recording must match the mode used for participant recordings. Mismatching modes silently under-reads small pupils.

---

## Installation

From PyPI:

```bash
pip install eyelink1000plus-pupil-size-to-mm
```

With uv:

```bash
uv add eyelink1000plus-pupil-size-to-mm
```

For a local checkout, install editable:

```bash
pip install -e .
# or
uv add --editable .
```

The `record` subcommand drives an EyeLink 1000 Plus through SR Research's `pylink` C bindings, which are not on PyPI. Install them separately:

```bash
uv pip install --extra-index-url https://pypi.sr-support.com sr-research-pylink
```

You also need the **EyeLink Developers Kit** (native C libraries) from <https://www.sr-research.com/support/thread-13.html>.

After install, the `eyelink1000plus-pupil-size-to-mm` command is available on your `PATH`. The `record` subcommand requires a connected EyeLink Host PC; `compute`, `convert`, and `export-setup` are pure post-processing and have no EyeLink-side dependencies.

---

## Step-by-step workflow

### 0. Prepare the artificial eye

Print a black filled circle of known diameter on plain paper with a laser printer. SR Research's FAQ uses 7–8 mm. Measure the printed dot with calipers to confirm the actual diameter — laser-printer rendering can be a few percent off, and the unit-to-mm constant inherits that error linearly.

Mount the paper on the head-rest at the same camera-to-eye distance you use for participants. Mount it at the lateral position of the eye you are calibrating (the constant is per-eye).

### 1. Describe your physical setup

Generate a one-time `setup.json` describing your monitor, camera geometry, and EyeLink hardware/protocol configuration:

```bash
eyelink1000plus-pupil-size-to-mm export-setup ./setup.json
```

Open `setup.json` and replace every value with your own:

```json
{
  "_comment": "Example setup ... Replace every value with your own monitor, camera, and hardware config before recording.",
  "output_dir": "./data",
  "screen_res": [1920, 1080],
  "screen_width_mm": 531.36,
  "screen_height_mm": 298.98,
  "screen_distance_top_mm": 905.0,
  "screen_distance_bottom_mm": 920.0,
  "camera_to_screen_distance_mm": 925.0,
  "artificial_pupil_diameter_mm": 7.0,
  "duration_s": 10.0,
  "el_configuration": "MTABLER",
  "camera_lens_focal_length_mm": 35,
  "sampling_rate_hz": 1000,
  "pupil_size_mode": "AREA"
}
```

Every key is required. Key meanings:

| Key | Meaning |
|---|---|
| `output_dir` | Where the EDF (and the rest of the recording artifacts) are written. |
| `screen_res`, `screen_width_mm`, `screen_height_mm` | Display resolution and physical size. |
| `screen_distance_top_mm`, `screen_distance_bottom_mm` | Eye-to-screen distance at the top and bottom edges (encodes screen tilt). |
| `camera_to_screen_distance_mm` | Distance from the EyeLink camera lens to the screen surface. |
| `artificial_pupil_diameter_mm` | Caliper-measured diameter of the printed dot. |
| `duration_s` | Length of the artificial-eye recording. |
| `el_configuration` | EyeLink physical configuration (e.g. `MTABLER` monocular desktop, `BTABLER` binocular desktop, `RTABLER` remote desktop). Must match your hardware. |
| `camera_lens_focal_length_mm` | EyeLink lens focal length, typically `25`, `35`, or `50`. |
| `sampling_rate_hz` | EyeLink sampling rate: `250`, `500`, `1000`, or `2000`. Must match the participant-recording rate. |
| `pupil_size_mode` | `AREA` (EyeLink default) or `DIAMETER`. Must match the Host PC's `pupil_size_diameter` setting used for participant recordings. |

`export-setup` refuses to overwrite an existing file, so editing `setup.json` after generating it is safe.

### 2. Record the artificial eye (left and right separately)

Pupil-only tracking is enabled at recording start; no `FINAL.INI` edits or Host-PC reboot needed. Settings revert when the connection closes.

Two CLI paths, **mutually exclusive**:

**A. Point at a setup JSON (recommended):**

```bash
eyelink1000plus-pupil-size-to-mm record --side L --setup ./setup.json
eyelink1000plus-pupil-size-to-mm record --side R --setup ./setup.json
```

**B. Pass every setup field as an individual flag (no JSON):**

```bash
eyelink1000plus-pupil-size-to-mm record --side L \
    --output-dir ./data \
    --screen-res 1920 1080 \
    --screen-width-mm 531.36 --screen-height-mm 298.98 \
    --screen-distance-top-mm 905 --screen-distance-bottom-mm 920 \
    --camera-to-screen-distance-mm 925 \
    --artificial-pupil-diameter-mm 7 \
    --duration-s 10 \
    --el-configuration MTABLER \
    --camera-lens-focal-length-mm 35 \
    --sampling-rate-hz 1000 \
    --pupil-size-mode AREA
```

On the Host PC's camera-setup screen, frame the artificial eye and confirm a stable pupil lock with no corneal reflection (the Host PC is already switched to PUPIL-only by the runtime commands), then exit setup to start the timed recording.

Two EDFs are produced: `<output_dir>/pupil_calib_7mm_left.edf` and `<output_dir>/pupil_calib_7mm_right.edf`. Override the stem with `--filename` if needed; pyelink prompts to replace/rename if the EDF already exists.

Each EDF is paired with a `<filename>.setup.json` sidecar containing the full setup, the eye flag, the tool version, and a UTC timestamp. `compute` reads the artificial-pupil diameter and recording mode from this sidecar.

### 3. Convert each EDF to JSON

[`syelink`](https://github.com/mh-salari/syelink) parses SR Research ASCII files into JSON:

```bash
syelink convert ./data/pupil_calib_7mm_left.edf  --json
syelink convert ./data/pupil_calib_7mm_right.edf --json
```

Any tool that produces a JSON with a `gaze_samples` list containing per-sample `left_pupil` / `right_pupil` fields will work — see the JSON schema notes below.

### 4. Compute the per-eye unit-to-mm constant

Run `compute` once per eye. Both runs merge into the same `pupil_units_per_mm.json` without overwriting each other. Calibration parameters come from the setup sidecar written by `record` (mutually exclusive with passing them explicitly):

```bash
eyelink1000plus-pupil-size-to-mm compute \
    --eye left \
    --input      ./data/pupil_calib_7mm_left.json \
    --setup-json ./data/pupil_calib_7mm_left.setup.json \
    --output     ./data/pupil_units_per_mm.json

eyelink1000plus-pupil-size-to-mm compute \
    --eye right \
    --input      ./data/pupil_calib_7mm_right.json \
    --setup-json ./data/pupil_calib_7mm_right.setup.json \
    --output     ./data/pupil_units_per_mm.json
```

If you don't have the sidecar (e.g., you produced the recording outside this tool), pass `--known-mm` and `--mode` instead:

```bash
eyelink1000plus-pupil-size-to-mm compute \
    --eye left --known-mm 7 --mode area \
    --input  ./data/pupil_calib_7mm_left.json \
    --output ./data/pupil_units_per_mm.json
```

`--setup-json` and (`--known-mm` + `--mode`) are mutually exclusive — pass exactly one path. The recording mode must match the Host PC's `pupil_size_diameter` setting at recording time (`area` for the EyeLink default; `diameter` if the Host PC has `pupil_size_diameter = YES`). If the two `compute` runs disagree on mode, the second run errors.

### 5. Apply the calibration to a participant recording

```bash
eyelink1000plus-pupil-size-to-mm convert \
    --input ./data/some_recording.json \
    --calibration ./data/pupil_units_per_mm.json
```

`convert` adds `left_pupil_mm` and `right_pupil_mm` to every sample (in place) for whichever eyes have raw pupil data. To restrict to a subset, pass `--eyes left_eye right_eye` (the default auto-detects from the sample data).

The same `pupil_units_per_mm.json` is reused for every recording made with the same physical setup; you only re-derive it if the camera-to-eye distance or the Host PC's `pupil_size_diameter` setting changes.

---

## CLI reference

```text
eyelink1000plus-pupil-size-to-mm export-setup OUTPUT
eyelink1000plus-pupil-size-to-mm record       --side {L,R} (--setup PATH | <every setup flag>) [--filename STEM] [--dummy]
eyelink1000plus-pupil-size-to-mm compute      --eye {left,right} --input JSON --output JSON (--setup-json PATH | --known-mm MM --mode {area,diameter})
eyelink1000plus-pupil-size-to-mm convert      --input JSON --calibration JSON [--eyes left_eye right_eye]
```

Each subcommand supports `--help` for the full option listing.

---

## Library API

For programmatic use, the three pure functions are re-exported from the package root:

```python
from pathlib import Path

from eyelink1000plus_pupil_size_to_mm import (
    augment_gaze_samples,
    load_calibration,
    units_to_mm,
)

mode, constants = load_calibration(
    Path("pupil_units_per_mm.json"),
    eyes=["left_eye", "right_eye"],
)

# One value:
mm = units_to_mm(value=1234.0, mode=mode, constant=constants["left_eye"])

# A whole sample list, in place; returns the count of samples that got a non-None mm value:
n = augment_gaze_samples(samples, eyes=["left_eye", "right_eye"], mode=mode, constants=constants)
```

The CLI subcommands are thin wrappers around three callables that mirror them: `eyelink1000plus_pupil_size_to_mm.record.record_artificial_eye`, `…compute.compute_for_eye`, and `…convert.convert_recording`.

For programmatic recording, `record_artificial_eye(eye_flag, setup, *, dummy=False, filename=None, extra_commands=())` takes a `setup` dict with the same keys as `setup.json` (no defaults — every key required). `extra_commands` is a tuple of additional Host-PC commands sent after the pupil-only commands and before Camera Setup, useful for lab-specific overrides such as `remote_camera_position`.

---

## Input/output JSON schema

`compute` and `convert` operate on JSON files with this minimal shape (additional fields are passed through untouched):

```json
{
  "gaze_samples": [
    {
      "left_pupil":  1234.0,
      "right_pupil": 1187.0
    }
  ]
}
```

`convert` augments each sample with `left_pupil_mm` and/or `right_pupil_mm`, in place. The output is written back to the same file. `convert` is idempotent — re-running on an already-augmented JSON recomputes the mm values from the original `*_pupil` fields.

The calibration JSON written by `compute` has this shape:

```json
{
  "mode": "area",
  "left_eye": {
    "constant": 11.2949,
    "calibration_file": "...",
    "known_diameter_mm": 7.0,
    "n_samples": 8123,
    "mean_units": 6256.4,
    "sd_units": 88.1
  },
  "right_eye": { "constant": 11.6730, "...": "..." }
}
```

---

## Reference

SR Research's procedural FAQ this tool implements is shipped under [`docs/`](docs/) for offline reference. The original is at <https://www.sr-research.com/support/printthread.php?tid=154>.

---

## Acknowledgments

This project has received funding from the European Union's Horizon Europe research and innovation funding program under grant agreement No 101072410, Eyes4ICU project.

<p align="center">
<img src="https://raw.githubusercontent.com/mh-salari/eyelink1000plus-pupil-size-to-mm/main/resources/Funded_by_EU_Eyes4ICU.png" alt="Funded by EU Eyes4ICU" width="500">
</p>